Introduction
• Simplicity: 4 static casters with self-locking function.
• Precision in an anesthesia ventilator with multiple ventilation modes
• Flexible configurations able to customize your requirements.
• International standard and advanced technology suitable for wide-range use.
• Compact interface and big screen give you a better operating experience.
• Over 5,000 units are installed in more than 200 countries.
Feature
• 10.4" TFT LCD screen displays the Ventilation parameters, Alarm information, and Oscillogram.
• High precision flowmeter, instantly know the fresh gas flow to your patient.
• Integrated breathing circuit design, ensure easy operating and keep tidy.
• Multiple working modes such as volume control and pressure limit, adapt to a wide range of patients.
• Vaporizer with temperature, pressure, flow compensation, and self-lock function, keep safety anytime.
• Multiple parameters monitoring interface, make every parameter clear, let users know the patient conditions in all aspects.
• Pressure-time, low-time loop oscillogram, and high precision ETCO2, O2 concentration show in real-time.
• Vital sign monitor and anesthetic gas monitor are optional.
• ETCO2 module and Anesthesia Gas Scavenging System (AGSS) are optional
Safety
• Three-level alarm system, visual and sound alarm information.
• With multiple types of alarm, reminder, and protection functions.
• Advanced power management control technology.
• Built-in backup battery provides the emergency power supply to the unit.
• Low O2 pressure alarm and N2O cut-off protection.
Specification
1. Ventilation mode
IPPV, A/C, PCV, SIMV,SIGH, MANUAL
2. Ventilator parameter range
Flowmeter
O2(0.1-10L/min)
N2O(0.1-10L/min)
Air(0.1-10L/min)
Rapid oxygen supply: 25L/min-75 L/min
Tidal volume (Vt): 0, 20mL-1500 ml
Frequence: 1 /min--100/min
I:E: 4:1-1:8
PEEP: 0cmH2O-30 cmH2O
Pressure triggering sensitivity (Ptr): -20cmH2O-0cmH2O(Based on PEEP)
Flow trigger sensitivity (Ftr): 0.5 L/min-30L/min
Pressure control(PC): 5cm H2O-60cmH2O
SIGH: 0 (off) 1/100-5/100
Apnea ventilation: OFF, 5s~60s
Pressure limit: 20 cmH2O-100 cmH2O
Monitoring parameter
1. Frequence: 0/min--100 /min
2. Tidal volume (Vt): 0mL -2000mL
3. MV: 0L/min-100 L/min
4. Oxygen concentration: 15%-100%
Oscillogram
P-T (pressure-time)
F-T (flow -time)
P-V Loop(pressure-volume loop)
Alarm and protection
1. The AC power failure alarm: Power failure or no connection
2. Low voltage alarm for back up battery: <11.3±0.3V
3. No tidal volume: ≤5mL within 6s
4. High oxygen concentration alarm: 19%~100%
5. Low oxygen concentration alarm: 18%~99%
6. High airway pressure alarm: 20cmH2O-100 cmH2O
7. Low airway pressure alarm: 0cmH2O-20cmH2O
8.High minute volume alarm: Adult (5L/min-20L/min)
Paed (1L/min-15L/min)
9. Low minute volume alarm: 0-10L/min
10. Continuous pressure alarm: (PEEP+1.5 kPa) over 16s
11. Suffocation warning: 5s-60s no spontaneous ventilation
12. The maximum limited pressure: <12.5kPa
13. Fan error: Show onscreen
14. Oxygen deficit: Show onscreen
Working condition
1. Gas source: O2,N2O,Air
2. Pressure: 280 kPa-600 kPa
3. Voltage: 100-240V
4. Power frequency: 50/60 Hz
Packing:
1. Wooden case packing size: L 920*W970*H1380mm
2. G.W.: 156KG
3. N.W.: 102KG
4. CBM: 1.23m2