2. We listen to our customers and integrate their insights into our machine development. Flexible configurations to suit customers' needs.
3. Countless feedback from customers gives good reviews.
1. Providing customers with high-quality and cost-effective anesthesia machines.
2. We listen to our customers and integrate their insights into our machine development. Flexible configurations to suit customers' needs.
3. Countless feedback from customers gives good reviews.
4. Real-time pressure-time,flow-time loop Oscillogram, and high precision, concentration detection function included.
1. 7" TFT LCD screen displays the Ventilation parameters, Alarm information, and Oscillogram.
2. High precision flow meter, instantly know the fresh gas flow to your patient.
3. Integrated breathing circuit design, ensure easy operating, and keep tidy.
4. Multiple working modes such as volume control and pressure limit, adapt to a wide range of patients.
5. Vaporizer with temperature, pressure, flow compensation, and self-lock function, keep safety anytime.
6. Multiple parameters monitoring interface, make every parameter clear, let users know the patient conditions in all aspects.
7. Pressure-time, and flow-time oscillograms show in real-time.
8. Vital sign monitor and Anesthetic gas monitor are optional.
1. Three-level alarm system, visual and sound alarm information.
2. With multiple types of alarm, reminder, and protection functions.
3. Advanced power management control technology.
4. Low O2 pressure alarm and N2O cut-off protection.
IPPV, A/C, SIMV,SIGH, MANUAL
Flowmeter
O2(0.1-10L/min)
N2O(0.1-10L/min)
Rapid oxygen supply: 25L/min-75 L/min
Tidal volume (Vt): 0, 20mL-1500 ml
Frequence: 1 /min--100/min
I:E: 2:1-1:6
Pressure triggering sensitivity (Ptr): -20cmH2O-0cmH2O (Based on PEEP)
Flow trigger sensitivny (Ftr): 0.5 L/min-30L/min
SIGH 0 (off) : 1/100-5/100
Apnea ventilation OFF: 5s~60s
Pressure limit: 20 cmH2O-100 cmH2O
The maximum limited pressure: <12.5kPa
Fan error: Show on screen
Oxygen deficit: Show on screen
The AC power failure alarm: Power failure or no connection
Low voltage alarm for backup battery: <11.3±0.3V
High oxygen concentration alarm: 19%~100%
Low oxygen concentration alarm: 18%~99%
High airway pressure alarm: 20cmH2O-100 cmH2O
Low airway pressure alarm: 0cmH2O-20cmH2O
High minute volume alarm:
Adult (5L/min-20L/min)
Paed (1L/min-15L/min)
No tidal volume: ≤5ml within 6s
Low minute volume alarm: 0-10L/min
Continuous pressure alarm: (PEEP+1.5 kPa) over 16s
Suffocation warning: 5s-60s no spontaneous ventilation
Monitor Parameters: Freq, Vt, MV, Paw
Graphical Display: P-T
Frequence: 0/min--100 /min
Tidal volume (Vt): 0mL -2000mL
MV: 0L/min-100 L/min
Oxygen concentration: 15%-100%
Working condition
Gas source: O2, N2O
Pressure: 280 kPa-600 kPa
Voltage: 100-240V
Power frequency: 50/60 Hz
Oscillogram
P-T (pressure-time)
F-T (flow -time)
Packing size(Wooden case ):
L740*W800*H1460mm,G.W.:98KG, N.W.: 66KG