1 | General Requirement |
1.1 | Anaesthesia machine with an integrated ventilator for adult to infants and integrated airway monitor for airway pressures and volume. |
1.2 | The machine should be suitable for low and minimal flow anesthesia application with compliance compensation of breathing ckt. Should have fresh gas compensation system. |
1.3 | The machine should have minimum 3 drawers |
1.4 | The anesthesia machine, inbuilt ventilator, vaporizer should be manufactured by same company. |
1.5 | The system should have upto 2 Hrs. battery backup |
2 | Gas delivery system |
2.1 | Should have pin index yokes for Oxygen, N2O & air one each besides separate connection for Central gas supply for Oxygen , N2O and Air. |
2.2 | Machine should provide electronic measurement and display flowmeter,with virtual double tube flowmeter for each gas on system screen display |
2.3 | Emergency Oxygen flush at 25 – 75 L/min bypassing the vaporizer. |
2.4 | In case of electricity and battery failure, manual ventilation, gas and agent delivery should be possible. |
3 | Vaporizer |
3.1 | Machine should have possibility to mount two quick mount type vaporizer for easy interchangeability, and safety with interlock facility. |
3.2 | Should be provided with Temperature / pressure/ flow compensated |
3.3 | Should be capable of connecting Vaporiser for Isofluorene / Sevoflourane / Halothene Vaporizer |
4 | Breathing System |
4.1 | Should have semi closed circle absorber system. |
4.2 | The system should have leak and compliance test (including patient hoses upto the Y piece). |
4.3 | All materials in contact with exhaled patient gases are autoclavable, except mechanical pressure meter and O2 cell |
4.4 | System should have built in heating system so as to provide warm gas to the patient. |
4.5 | Carbon dioxide absorbent canister should be detachable intra-operatively with BYPASS function. (OPTIONAL) |
4.6 | Should have an external fresh gas outlet (ACGO) for connecting Magill or Bain’s circuit (OPTIONAL) |
5 | Anaesthesia Ventilator |
5.1 | The system should have inbuilt ventilator with Pneumatically driven and electronically controlled |
5.2 | Should not require changing of bellows for adult & infants. |
5.3 | ventilation modes: Volume Controlled(VCV), Pressure Controlled (PCV), Pressure Controlled Ventilation –Volume Guaranteed (PCV-VG), Synchronized Intermittent Mandatory Ventilation (SIMV-VC and SIMV-PC), Pressure Support with Apnea Backup (PS), MANUAL; |
5.4 | Tidal Volume: 20ml -1500 ml |
5.5 | Electronic PEEP: OFF, 3 to 30 cmH2O. |
5.6 | Breathing Frequency : 4 to 99 BPM |
5.7 | I:E ratio: 4:1 to 1:8 |
5.8 | The pressure range: 5 to 70 cmH2O |
5.9 | The Inspiratory Slope Time (Tslope) range: 0-2S |
6 | Airway monitoring |
6.1 | Should have integrated touch screen of at least 12” |
6.2 | Integrated monitor for electronic monitoring and display of following parameters : |
6.3 | Tidal Volume |
6.4 | Minute volume |
6.5 | PEEP, Peak & Mean and Plateau airway pressure |
6.6 | Frequency |
6.7 | Waveform display for V-T,F-T,P-T synchronously |
6.8 | Loop display for pressure-volume loop and flow -volume loop; |
6.9 | FiO2 Monitoring |
6.10 | EtCO2 Monitoring (OPTIONAL) |
6.11 | Anaesthesia Gas Monitoring with 5AA AUTO ID and MAC value(OPTIONAL) |
7 | Alarm limits & alarms |
7.1 | Adjustable high / low limits with audio and visual alarms for the following : |
7.2 | Minute volume, |
7.3 | Airway pressure |
7.4 | Insp oxygen concentration, |
7.5 | Audio power supply fail alarm, |
7.6 | Anesthesia ventialtor should have 240 hrs trend data and 72 hrs chart of monitoring parameters. |
7.7 | Should stores 500 alarm events |
8 | Scope of supply |
8.1 | Anaesthesia machine main unit |
8.2 | Working table and auxiliary folding writing surfacetable |
8.3 | 4 Electric Sockets |
8.4 | Pin Index yokes for O2 , N2O & Air. |
8.5 | Isofluorene vaporizer. |
8.6 | Adult & Pediatric silicon breathing accessories.(OPTIONAL) |
8.7 | Anesthetic mask size – Adult & child (OPTIONAL) |
8.9 | Central gas supply hoses (Color coded) |
9 | Operating Environment |
9.1 | The system offered shall be designed to be stored and to operate normally under the conditions of the purchaser's country. The condition include Power Supply, Climate, Temperature Humidity, etc |
9.2 | Power supply: 100 – 240 V AC, 50/60 Hz fitted with appropriate plug. The power cable must be at least 5m in length. |
10 | Standards and Safety Requirements |
10.1 | CE approved |