Display mode | High-definition 5.7 LCD screen display |
Ventilation mode | IPPV, SIPPV, SIMV, MANUAL |
Ventilation function | PEEP, SIGH, IRV |
Ventilation parameters | Tidal volume: 50-1500ml Rate: 2-99bpm SIMV rate: 2-20bpm I:E: 2:1-1: 8 Inspiratory trigger pressure: -10-20cmH2O PEEP: 2-20cmH2O Pressure limitation: 10-60cmH2O SIGH: 1.5 times the inspiratory time/100bpm |
Parameters for ventilation monitoring | Tidal volume, Minute Ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure-time waveform, flow rate-time waveform, PEEP, inspiratory trigger pressure, average pressure, oxygen concentration, battery capacity |
Security Alert System | Oxygen Concentration alarm, Airway pressure alarm, Minute ventilation volume alarm, Tidal volume, Sustained high-pressure alarm, Suffocation alarm, Power alarm, Intubation off the alarm, battery voltage alarm |
Oxygen Concentration alarm | Upper limit setting range: 21%-99% Low limit setting range: 18%-80% |
Airway pressure alarm | Upper limit setting range: 5-60cmH2O Low limit setting range: 0-50cmH2O |
Minute ventilation volume alarm | Upper limit setting range: 1~20L/min Low limit setting range: 0~19L/min |
Tidal volume | Upper limit setting range: 100-1500ml Low limit setting range: 0-1400ml |
Sustained high-pressure alarm | It will give an alarm when stress has consistently been higher than 25cmH2O |
Power | AC 220V 50Hz (UPS, Storage battery) |
Function | Mechanical Arms |
Optional Items | End-tidal CO2 monitoring Specific Interface of waste anesthesia gas |